Resonance R07a Screening Portable Audiometer
The Resonance R07A is a small Compact Occupational Health Audiometer, it connects to the a PC or Laptop using a USB Cable. For Occupational Health Audiometry, it is completely automatic, using the Hughson Westlake method for automatic audiometry. Tests are usually completed within a matter of minutes. The Resonance R07A is compatible with our Occupational Health Software.
GENERAL SPECIFICATIONS:
DIMENSIONS AND WEIGHT
- L x W x H: 240 x 190 x 30 mm
- Net Weight: 600 gr. without battery
- Body material: Plastic cabinet
TEST TYPES
- Pure Tone Audiometry, Automatic Audiometry Patient controlled, according to ISO 8253-1 and standard Hughson Westlake procedure (selectable on control panel).
- Audiometer Class 4 according to IEC 60645-1
TALK FORWARD
- Built-in talk forward microphone 50-110 dBSPL (adjustable on control panel)
STORE FUNCTION
- Dedicated Store button that enables to store measurements into the internal memory (up to 50 patients). The patient will be identified by an ID. number.
- Stored data can be viewed on the built-in display or transferred to the PC using MDS software.
PC INTERFACE
- Resonance MDS Management Data Suite, with reporting and printing features.
AUDIOMETRY OPERATING SPECIFICATIONS
RANGE
Frequency range (AC):
- 125 – 8000 Hz (DD45)
- 125 – 8000 Hz (Insert Phones)
INPUTS
- Tone
- Warble Tone +/- 5%, 5 Hz (true sine wave frequency modulation)
OUTPUTS
- Left and Right
TONE PRESENTATION
- Manual or Reverse (selected on Setup Menu)
- Pulsed at 0.5,1 or 2 Hz – (selected on Setup Menu) or Continuous.
FREQUENCIES AND INTENSITIES WITH DD45
| Frequency (Hz) | Intensity (dB HL) |
|---|---|
| 125 | 65 |
| 250 | 85 |
| 500 | 100 |
| 750 | 100 |
| 1000 | 100 |
| 1500 | 100 |
| 2000 | 100 |
| 3000 | 100 |
| 4000 | 100 |
| 6000 | 100 |
| 8000 | 85 |
All the frequencies can be enabled or disabled by the Setup menu except 1000 Hz.
LEVEL STEPS
- 1, 2 or 5 dB (selected on Setup Menu)
DISTORTION
- 0.3% typical
- below 1% at full intensity
RISE/FALL TIMES
- 40 msec.
POWER:
BATTERIES
- 4 AA size
- Automatic battery off switching (battery save mode activated after 1, 2 , 5 minutes on non-activity or never) Battery status indicator on the display header. Protection against test execution with a critical battery level.
EXTERNAL POWER SUPPLY OPTIONAL
- Dedicated Medical approved power supply 100 – 240
- VAC 50/60 Hz, 0.3 A (input) +5 VDC, 2 A (output)
ENVIRONMENTAL:
OPERATING ENVIRONMENT
- Storage: -20° C up to +50° C
- Operating: +15° C up to +35° C
- Humidity: up to 90%, (non-condensing)
- Ambient pressure: from 700 hPa up to 1060 hPa
QUALITY SYSTEM
- Manufactured, designed, developed and marketed under an ISO 13485, ISO 9001 certified quality system. Medical CE marks and FDA approval.
COMPLIANCE / REGULATORY STANDARDS
- Meets or exceeds IEC 60645-1 type 4 and ANSI 3.6
- Safety standard: IEC 60601-1 3rd edition, Class II, type BF
- EMC: IEC 60601-1-2
STANDARD ACCESSORIES
- HB-7 Audiometric headband with DD45 transducers
- Patient pushbutton
- 100 Audiogram charts
- Blue and Red Pen
- Carrying bag
- Multilingual CE Operation Manual
- MDS Software
OPTIONALS
- Insert phones IP30
- HB-7 Audiometric headband with TDH39 transducers
- Medical CE approved external power supply
- ADC Audiocups Noise Excluders